If you have been in search of Analyzer Blood Gas products that are listed with the NSN 6630013648555 we have you covered here on Speedy NSN. You have reached the RFQ page for part number 4300M-W-PROVISIONING a Mallinckrodt Sensor Systems Inc item that is readily available for procurement from our website at any time. Whether you wish to purchase this part or simply want to receive more information for your comparisons, be sure to fill out and complete the form provided on this page with as much detail as you can regarding your needs and restrictions. Within 15 minutes of us receiving and reviewing a completed form, a member of our staff will reach out to you to present a customized quote that keeps your part requirements in mind.
Note : We Will Not Share Your Information To Any Third Parties.
To ensure that customers only procure the exact items they are looking for, we list products with as much information as possible. For example, this particular Mallinckrodt Sensor Systems Inc part was produced under the CAGE Code 0CNN1, ], a unique identifier assigned to manufacturers who have been approved to supply government entities. Additionally, with the NSN 6630-01-364-8555 this product also belongs to the FSC 6630 Chemical Analysis Instruments. If you ever have any questions regarding our offerings or services, be sure to call or email us at your convenience. Get started on the purchasing process today so we can fulfill all your operational needs as your strategic sourcing partner.
MRC | Criteria | Characteristic |
---|---|---|
DPTH | Depth | 18.300 INCHES |
ELEC | Voltage In Volts | 120.0 AND 230.0 |
MRC | Decoded Requirement | Clear Text Reply |
CBBL | Features Provided | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
CXCY | Part Name Assigned By Controlling Agency | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
ACDC | Current Type | AC |
CYFV | Method Of Analysis | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
ACZB | Frequency Rating | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
CYFY | Analysis Test/Measurement Range | PO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
CYHM | Test Sample/Specimen Type | WHOLE BLOOD |
ABRN | Height | 8.400 INCHES |
WDTH | Width | 9.100 INCHES |
AQXY | Test Type For Which Designed | BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
CYFZ | Test Sample/Specimen Size | 0.5 ML |
FEAT | Special Features | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |